Overview
Choose tip features around contamination risk and liquid behavior
Not all pipette tips perform the same function. Tip design influences contamination control, liquid recovery, sealing, and reproducibility—especially when you work with small volumes or challenging samples.
When choosing pipette tips, start with the application rather than simply matching the nominal volume. The most important decisions are whether you need an aerosol barrier and whether the liquid benefits from a low-retention surface.
Filter vs. Non-Filter Tips
Non-Filter Tips
Non-filter tips contain no internal aerosol barrier. They are appropriate for many routine applications where contamination risk is low, including:
- Aqueous buffer transfers
- Routine reagent preparation
- General sample dilution
- Non-sensitive laboratory workflows
Because there is no filter occupying space inside the tip, non-filter formats can also provide maximum usable tip capacity.
Filter Tips
Filter pipette tips contain a hydrophobic barrier between the liquid and pipette cone. The barrier is designed to reduce aerosols and liquid droplets reaching the pipette interior, helping protect the instrument and reduce the risk of carryover into subsequent samples. Filter construction and pore architecture vary by product rather than following one universal pore-size specification.
Use filter pipette tips when aerosol contamination could compromise the workflow, particularly for:
- PCR and qPCR setup
- RNA and DNA handling
- Cell culture
- Infectious or hazardous samples
- Low-copy-number molecular assays
A filter tip does not replace good pipetting practice. Change tips between samples and follow your laboratory's contamination-control procedures.
Standard vs. Low Retention Tips
Standard Tips
Standard polypropylene tips are suitable for most routine aqueous solutions. For water-like buffers and moderate transfer volumes, a standard tip is often the most economical choice.
Low Retention Tips
Low retention tips use surface properties designed to reduce liquid wetting and the film that remains inside the tip after dispensing.
They are especially useful for:
- Glycerol-containing solutions
- Serum and concentrated protein solutions
- Antibodies and other valuable reagents
- Detergent-containing solutions such as Tween-type or Triton-type surfactants
- Low-surface-tension liquids
- Small-volume transfers where residual liquid represents a meaningful percentage of the intended volume
Published comparative studies show that residual liquid becomes particularly important in small-volume tips and with detergent- or glycerol-containing solutions because the surface-to-volume ratio is less favorable.
The benefit of low retention tips is therefore not that they make every transfer inherently more accurate. They reduce one specific source of error—liquid remaining on the internal tip surface—which can improve sample recovery and consistency when retention is significant.
Both Universal Pipette Tips and LTS-Compatible Pipette Tips can be offered in standard and low-retention configurations depending on the required workflow.
Other Important Tip Features
Sterility
Choose sterile tips for workflows where microbial contamination must be controlled, including many cell culture applications.
Sterility and molecular purity are separate specifications. A sterile product should not automatically be assumed to meet every nuclease- or PCR-related requirement.
DNase/RNase-Free
For RNA and DNA workflows, verify that tips are certified to meet the required DNase- and RNase-free specifications.
Extended Length
Extended-length tips help you access deep-well plates, reagent vessels, or tubes while reducing the likelihood that the pipette shaft contacts the vessel opening.
Fine-Point Tips
Fine or narrow distal geometries can improve access to small wells and support controlled dispensing in low-volume applications.
Automation Tips
Automation Tips are manufactured for robotic pickup, sealing, positioning, and ejection. They require tighter dimensional consistency than typical manual tips and must be matched to the specific liquid-handler platform.
Decision Guide for Choosing Pipette Tips
| Application | Recommended Tip Type |
|---|---|
| Routine aqueous buffers | Standard, non-filter |
| Cell culture media | Sterile; filter when aerosol protection is required |
| qPCR / PCR setup | Filter, contamination-controlled |
| RNA/DNA workflows | Filter, DNase/RNase-free |
| Protein or antibody solutions | Low retention |
| Viscous liquids such as glycerol or serum | Low retention |
| Detergent-containing solutions | Low retention |
| Precious or limited samples | Low retention |
| Automated liquid handling | Validated Automation Tips |
| Infectious or aerosol-generating samples | Filter tips |
| Deep-well plates | Compatible extended-length tips |
When choosing pipette tips, several features can be combined. A PCR workflow, for example, may require a sterile or molecular-grade filter pipette tip with low-retention characteristics.
Cost Considerations
Filter and low-retention formats generally cost more than basic standard tips, so there is little benefit in specifying premium features where the application does not require them.
For routine buffer preparation, standard non-filter tips are often sufficient.
For contamination-sensitive molecular workflows, the additional cost of filter tips is easier to justify because one contamination event can invalidate an entire experiment.
Likewise, low retention tips are most valuable when you work with expensive samples, difficult liquid properties, or volumes small enough that residual liquid materially affects recovery.
Automation Tips typically carry additional manufacturing and packaging requirements. In robotic workflows, however, consistent pickup, sealing, and ejection can be more important than minimizing tip cost alone because a failed tip can interrupt an automated run.
FAQ
Common questions
Do filter tips affect pipetting accuracy?
A properly designed filter tip should support accurate pipetting within its validated volume range. The filter reduces internal air volume, and technique may need to be optimized for difficult liquids. Use smooth aspiration and validate critical quantitative workflows.
When should I use low retention vs. standard tips?
Use low-retention tips when handling viscous, detergent-containing, low-surface-tension, or valuable samples where residual liquid matters. Standard tips are appropriate for most routine aqueous transfers.
Can I reuse pipette tips?
Disposable pipette tips are generally intended for single-use workflows. Reuse increases the risk of cross-contamination, residual sample carryover, and changes in surface behavior. Only reuse a tip when your laboratory has a specifically validated procedure that permits it.
Products
Products used in this workflow
LTS-Compatible Pipette Tips
Standard and low-retention configurations.